Calibration of the Internal Weight (TW Only)
4
CALIBRATION
7
Take the calibration weight off the
pan.
Open the glass door of the windbreak, take the
weight off the pan, and close the glass door.
*
If "BUSY" is displayed…
There is something placed on the pan.
When this item is taken off the pan,
internal weight calibration will start
automatically.
To cancel internal weight calibration,
press
*
If "ERR H" is displayed…
See
.
*
If "ERR C" is displayed…
The internal weight has not been
calibrated for one of the following
reasons.
There is too large a discrepancy
between the zero point of the
balance and the sensitivity.
A container has been placed on the
pan.
The pan is not on the balance.
There is too large a discrepancy in
the value of the internal weight.
Press
and repeat the operation
from the beginning.
"END" is displayed, then span calibration using
the internal weight starts.
When span calibration using the internal weight
ends, the balance returns to the weighing mode.
^
"Span Calibration and Adjustment"
Caution
Instructions
If calibration doesn't end normally and the balance stops, do not
move it nor leave it as it is.
Moving the balance in such a condition may cause failure because the internal weight is
not held correctly.
Before moving the balance, be sure to turn the power on and start it up correctly (so that
the internal weight is correctly held).
!P.C!A!L!3
!P.C!A!L!2
!P.C!A!L!1
!P.C!A!L!0
! !W!A!I!T
!E!N!D
!P.C!A!L!3
!P.C!A!L!2
!P.C!A!L!1
!P.C!A!L!0
! !W!A!I!T
!E!N!D
* This may not
be displayed.
Содержание TW223L
Страница 101: ...BEFORE WEIGHING MAINTENANCE TROUBLESHOOTING FOR YOUR INFORMATION USING THE BALANCE USING MORE CONVENIENTLY 101 MEMO...
Страница 157: ...157 MEMO...
Страница 158: ......