Span Calibration and Adjustment
4
CALIBRATION
2
Enter the calibration weight value.
If necessary change the weight value displayed
to match the weight that will be used for
calibration. If there is no need to change it,
proceed to step
3
.
(Enter a weight value if necessary.)
^
"Entering Numerical Values", page 49
For details on the range of weight values that
can be used, see
With models that don't feature the windbreak
3
Place the calibration weight on the
pan.
Wait until the flashing weight value display
changes to a flashing zero.
4
Take the calibration weight off the
pan.
"END" will be displayed and the balance will
return to the weighing mode.
*
If "ERR C" is displayed…
Calibration has failed for one of the
reasons given below.
There is too large a discrepancy
between the zero point of the
balance and the sensitivity.
A container has been placed on the
pan.
The pan is not on the balance.
The wrong weight has been placed
on the pan.
No operation has been performed
within 60 seconds of the flashing
weight value or zero display.
Press
(
with the TXB series)
and repeat the operation from the
beginning.
!3!0!0.0!0!0
!2!9!9.9!9!8
/
/
//
/
/
/
/
/
/
/
/
/ /
/
/
/
/
/ /
/
/
/
//
//
//
/
/
//
//
/
/
/ /
/
/ /
/ / /
!3!0!0.0!0!0
!2!9!9.9!9!8
/
/
//
/
/
/
/
/
/
/
/
/ /
/
/
/
/
/ /
/
/
/
//
//
//
/
/
//
//
/
/
/ /
/
/ /
/ / /
! ! !0.0!0!0
/ / / /
/
/
/
//
/
/
//
/
/
//
/
/
/ /
/ / /
! ! !0.0!0!0
/ / / /
/
/
/
//
/
/
//
/
/
//
/
/
/ /
/ / /
!E!N!D