134
11
MAINTENANCE
In order to establish and maintain the performance of the balance, weights must be used to accurately
adjust the balance's scale, and to check its adjustment.
With the TW/TX/TXB series balances, weights are used when performing
calibration
and inspections
in the environment in which the balance is actually used. The weights
should be prepared in advance and managed properly.
Types of Weight and Their Selection
There are many types of weights.
Select the appropriate weights for the specifications of your balance by referring to the following
table.
Selecting the class of weight
As the main form of classification, weights are normally divided into classes according to their
degree of accuracy.
Select the most appropriate class of weights to be used for calibration and inspection of the
balance, based on the type of the balance.
The table below shows the classes of weight and the applicable balances.
Class of Weight
Applicable Type of Balance
Minimum Indication
Resolution*
Common Name
E2
Less than 1 mg
Around 1/1,000,000 or
better
Analytical balances
F1
1 mg or greater
Around 1/100,000 or
better
Toploading balances
F2
1 mg or greater
Around 1/100,000 or
lower
Toploading balances
M1
10 mg or greater
Around 1/10,000 or
better
Scales, etc.
* "Resolution" means: minimum indication / weighing capacity
About Weights