General Setup
Commands
#iiDateTime=mmddyyyyhhmmss
Set real-time clock month, day, year, hour,
minute, second. If you enter a date before
January 1, 2000 or after January 1, 2100,
39
plus
-IM resets date to January 1, 2000.
#iiOutputExecutedTag=x
x=Y
: Display XML Executing and
Executed tags. Executed tag displays at
end of each command response;
Executing tag displays one or more times
if 39
plus
-IM command response requires
more time.
Default.
x=N
: Do not. Note that tags always output
when communicating via USB, regardless
of this setting.
Note:
If the 39
plus
-IM is logging or waiting
to start logging (
#iiStartLater
has
been sent), it will not allow the user to
reset date or time.
Example:
Set current date and time for 39
plus
-IM with ID=03 to
10 July 2016 12:23:59 (user input in bold).
#03datetime=07102016122358
Example:
Set 39
plus
-IM with ID=03 to output Executed and Executing
tags (user input in bold).
S>outputexecutedtag=y
<Executed/>
S>
getcd
. . .
(
GetCD
response)
<Executed/>
S>