Manual revision 001
Section 4: Deploying and Operating SBE 39
plus
-IM
SBE 39
plus
-IM
51
Status
Commands (
continued
)
#iiDS
Display status and setup parameters.
List below includes, where applicable,
command used to modify parameter.
•
firmware version, serial number, date
and time [
#iiDateTime=
]
•
Voltages – main battery pack voltage
and back-up cell voltage
•
sampling status (logging, waiting to
start at . . ., or not logging, etc.), and
(if applicable) reason that logging
stopped (such as received stop
command, low battery pack power,
etc.)
•
Sample interval time
[
#iiSampleInterval=
]
•
Number of samples in memory and
available sample space in memory
•
Configuration (temperature only or
temperature and pressure); factory set
•
Output data format
[
#iiOutputFormat=
]
•
Output temperature
[
#iiOutputTemp=
]? Temperature
units [
#iiSetTempUnits=
]
•
Output pressure [
#iiOutputPress=
]?
Pressure units [
#iiSetPressUnits=
].
Only appears if 39
plus
-IM includes
pressure sensor.
•
Transmit sample number with
!iiData
,
Dataii
, or
!iiGetReply
output
[
#iiTxSampleNum=
]?
•
Current temperature (always in °C,
regardless of setting for
#iiSetTempUnits=
)
Example:
39
plus
-IM with ID=03
(user input in bold, command used to modify parameter in parentheses).
#03ds
SBE 39plus-IM V 4.4.0 SERIAL NO. 03907842 28 Apr 2016 21:26:15
[
#iiDateTime=
,
#iiFormat=
]
battery voltage = 7.12, back-up voltage = 3.30
not logging: received stop command
sample interval = 10 seconds
[
#iiSampleInterval=
]
samplenumber = 25815, free = 5566590
[reset with
#iiInitLogging=
]
configuration = temperature and pressure
data format = converted engineering
[
#iiOutputFormat=
]
output temperature, Celsius
[
#iiOutputTemp=
,
#iiSetTempUnits=
]
output pressure, Decibar
[
#iiOutputPress=
,
#iiSetPressUnits=
]
output sample number
[
#iiTxSampleNum=
]
temperature = 24.12 deg C
Notes:
•
The
#iiDS
response contains similar
information as the combined
responses from
#iiGetSD
and
#iiGetCD
, but in a different format.
•
Lines describing what parameters to
output (temperature, pressure, sample
number) only appear if they are
enabled and
#iiOutputFormat=1
or
2
.
Raw output (
#iiOutputFormat=0
) is
not affected by enabling / disabling
parameter outputs.
•
The
#iiDS
response is also affected
by the
#iiLegacy=
command.
See the Output Format Setup
Commands below.
•
If the 39
plus
-IM is logging data and
the battery pack voltage is less than
5 volts for ten consecutive scans, the
39
plus
-IM halts logging and sets the
logging status to low battery.
•
Replace the back-up cell if it drops
below 2.5 V.