Power and Communications Test – Using USB Communications
Test Setup
1.
Connect to Internal USB connector:
A.
Wipe the outside of the titanium end cap and housing dry, being
careful to remove any water at the seam between them.
B.
Using a wrench on the end cap’s wrench flats, unscrew the end cap.
C.
Pull the end cap and attached electronics out of the housing.
D.
Remove any water from the end cap O-rings and mating surfaces
inside the housing with a lint-free cloth or tissue.
E.
Connect the USB cable to the USB connector below the battery pack.
F.
Connect the USB cable to your computer’s USB port.
Test
1.
Double click on
SeatermV2.exe
. The main screen looks like this:
SeatermV2 is a
launcher
, and launches the appropriate terminal program
for the selected instrument.
Notes:
•
Sea-Bird ships the 39
plus
-IM with
four AA lithium cells installed. See
Replacing AA Cells and Desiccant
in
Section 5: Routine Maintenance
and Calibration
to replace the cells.
•
The 39
plus
-IM automatically stops
autonomous sampling when you
connect via USB.
Note:
See SeatermV2’s Help files.
USB cable
Wrench flat (both sides)
Shown with external
thermistor; detail similar
for 39
plus
-IM with
embedded thermistor
39
plus
-IM with external thermistor;
detail similar 39
plus
-IM with embedded thermistor