3 .2 MEASURES FOR YOUR OPERATIONAl SAFETY
Make sure the
EasyPREP
BOD-300
is placed on a surface capable of supporting 200 pounds (100 kg).
EasyPREP
BOD-300
requires a clearance of five inches (12.7cm) on all sides.
If moving the system, always use 2 people and do not lift by the Kydex
®
or Corian
®
panels. Lift by the bottom aluminum frame instead.
Keep hands and fingers free of the system when it is moving to reduce the risk of pinching.
Always plug the
EasyPREP
BOD-300
in a properly grounded three-prong electrical outlet (100-240V receptacle).
Use a stabilized constant voltage AC power supply, with a voltage /-5% of the specified level.
Always use the provided mains power cord or one with the same manufacturer’s specifications.
If the
BOD-300
is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.