“Speed X”: Allow speed adjustment on X axis (1-100-- 1 is the slowest and 100 is the fastest. The default factory setting is 100).
•
“Speed Y”: Allow speed adjustment on Y axis (1-100-- 1 is the slowest and 100 is the fastest. The default factory setting is 100).
•
“Speed Z”: Allow speed adjustment on Z axis (1-100-- 1 is the slowest and 100 is the fastest. The default factory setting is 100).
•
“
go
” : Go to selected bottle position, use the drop down menu to select the position. A 00, B 00, and C 00 are coordinates for the wash
station.
6.5.2 Rack Parameters
The Rack Parameters section is used to configure and adjust the positioning of the instrument using X, Y and Z parameters. The racks are
configured using bottle 1 (back-left corner) of each rack (Figure 23).
The racks are configured using bottle 1 (back-left corner) of each rack (Figure 28).
Figure 23. Rack set parameters.
“x0”, “y0”
:
Coordinates of the first bottle within the rack.
“modiFy”
: Redefines the X0 and Y0 of a highlighted rack using the values placed in X0 and Y0, respectively.
“choose A rAck”
: Allows for different sized bottles to be selected and added to the platform.
“edit rAck”
: Allows for further modification of the racks beyond X and Y positioning.
Once the bottle size is defined as either 300, 165 or 65 ml, the left-back corner tube/bottle is defined by the X0 and the Y0. It is required
that the racks parameters are saved once all the required racks have been modified. The wash station parameters can be modified in the
Instrument Memory Parameters.