Version 1 .0
26 |
“Z cAP/de-cAP”
: Depth Decapper Accessory uses to decap bottles
“y correction”
: Offset parameter to account for deviation in height in the Y direction
6.5.3
Calibration of the X, Y and Z positioning for Autofill Sensor(s)
The Autofill sensor can be controlled in the Y and Z by the user. To ensure that the sensor enters the bottle smoothly and is not in contact with
the side of the bottle, the X and Y must be calibrated. The Memory can be changed by altering an individual cell and pressing enter. The cell
should turn yellow, The Memory parameters must then be updated.
The Memory parameters must then be Written/Reload, which will turn red as it is rewriting. The
BOD-300
must then be restarted.
The X coordinate of the probe is controlled by line 5 in the
EasyPREP
Memory Parameters, as shown in (Figure 26). Alteration of the
highlighted value will result in a shift in the x-axis of the Autofill sensor and the pumps. Increasing this value will result a shift to the left and
decreasing the value will result in a shift to the right.
Figure 26.
The Z coordinate of the Auto Fill Sensor is changed in ZAP as seen in (Figure 25). Z auto fill represents the Z value which the Autofill sensor
is lowered into the bottle when dispensing liquid. Increasing the Z auto fill will decrease the fill line of the bottles in question and decreasing Z
auto fill will increase the fill line of the bottles.
The y- coordinate can be controlled by line 79 in the
EasyPREP
Memory Parameters, and is highlighted below in (Figure 27). This can be a
positive or negative value. Alteration of this parameter will bring sensor forward by increasing the parameter and can move the sensor toward
the rear by decreasing the parameter.