Version 1 .0
| 21
•
Press "
run PumP
" and collect the dispensed water in a pre-weighed bottle.
•
Weigh the dispensed water, convert to volume and type the value in "Measure".
•
Press "
cAlibrAte
", the new pump flow rate will be reset under "Flow Rate".
Only press Calibrate ONCE.
6.4.3.3 Waste pump
Set timing for washing station.
•
"
Fill
": Set the number of seconds to fill the washing/waste station.
6.4.4 Priming/Cleaning
Used to prime the tubing (fill and remove any air bubbles prior to sample analysis) or clean the tubing (fill the tubing with a cleaning solution).
•
"
inhibitor/seed PumP
": Set volume to prime/clean the inhibitor pump.
•
"
seed PumP
": Set volume to prime/clean the Seed Pump
•
"
dilution WAter PumP
": Set volume to prime/clean the dilution water tubing.
Figure 20. Priming and cleaning
6.4.5 ProODO Probe
•
"
bArometric Pressure
": Reads the barometric pressure.
•
"
temPerAture
": Reads the sample temperature.
•
"
dissolved oxygen
": Reads the dissolved oxygen concentration (mg/L O
2
).
•
"
cAlibrAtion
": Calibrates the DO probe by one-point calibration in water-saturated air.
•
"
neW sensor cAP
": Input new sensor cap parameters (see section 8.2, page 34).
•
"
stirrer
": Manually start and stop the probe stirrer.
NOTE