Figure 21. Proodo probe.
6.4.5 Calibration of the BOD probe
Calibrate the DO probe by one-point calibration in water-saturated air. When calibrating, the probe lowers into position A1. The probe should be
sealing the opening to the bottle. After reaching 100% water-saturation in the air of the bottle (about 10 minutes), new calibration information
is obtained.
•
Fill a BOD bottle with 40 ml clean water and place it in position A1.
•
Press "Calibrate". When prompted for a calibration time, type "10" and then press Enter.
•
New calibration information is obtained and saved.
Though drift is minimal over time, to ensure the best precision, it is recommended to calibrate at the start of each day.