STATIM 7000 Service Manual – Cover Plate
7
3. Rear Chassis Cover Plate
Accessing the components behind the armature will require the removal of the
rear chassis cover plate. To remove this plate, follow these steps.
1.
Power
OFF
the unit, unplug it and remove the cover. See Cover. NOTE:
The steam generator may be hot if the unit has been operating.
2.
Disconnect the fan connectors.
3.
Disconnect the inlet and outlet tubes on the biological air filter.
4.
Remove the three screws across
the top and the three screws
across the bottom of the plate.
5.
Lift the plate to remove.
6.
Reverse instructions to replace.