STATIM 7000 Service Manual – Calibration
61
continue (default password is: Unwrapped, Wrapped, R&P and Stop keys pressed in
this order). The service technician can change this password. In case the changed
password is lost a backdoor password can be used: Unwrapped, Wrapped,
Unwrapped, Wrapped in this order.
Keypad functions at this time
:
Unwrapped Key:
Select next item in the menu.
Wrapped Key:
Select previous item in the menu.
Rubber and Plastic Key:
Enter current selection
Toggle using keypad through the menu selection to reach Calibration option and
press R&P key.
5. Confirm that the display appears similar to the example above.
CHAMBER CALIBRATION:
(Voltage reading and Chamber thermocouple)
6. Press the Unwrapped cycle button (first from the left) and after that press START
button to activate a chamber thermocouple calibration cycle. The calibration cycle will
start by entering the heating up phase and the screen will change for the
Voltage
Reading calibration
(see screen below):
Screen Representation
VVV = Voltage measured by unit
CCC = Voltage calibration offset.
Keypad functions at this time
Unwrapped
Increase current field
Wrapped
Decrease current field
Rubber and Plastics
Select and return to main menu
Stop
Exit, without saving, to normal mode of operation
VVV CVO TTT CO
DDD
Chamber
thermocouple
hexadecimal offset
Chamber
temperature in ºC
Difference between
Va l idation TC and chamber
temperatures in ºC
Va l idation thermocouple
hexadecimal offset
compared
with chamber reading
Va l idation
thermocouple
reading
Voltage Calibration
V=VVV VCAL=CCC