STATIM 7000 Service Manual –
Chamber Thermocouples
31
13.3 Validation Thermocouple
1. Power
OFF
the unit, unplug it and remove the cover. See Cover. NOTE:
The steam generator may be hot if the unit has been operating.
2. Disconnect the fan connectors.
3. Disconnect the inlet and outlet tubes on the biological air filter.
4. Remove the rear chassis cover plate by removing the 3 screws across the
top and 3 screws across the bottom, and lifting the cover plate.
5. Remove Armature. See Armature.
6. Loosen smaller top nut to remove
validation thermocouple.