STATIM 7000 Service Manual –
Chamber Thermocouples
27
13. Thermocouples
13.1 Auxiliary Heater Thermocouple (Inlet)
1. Power
OFF
the unit, unplug it and remove the cover. See Cover. NOTE:
The steam generator may be hot if the unit has been operating.
2. Disconnect the fan connectors.
3. Disconnect the inlet and outlet tubes on the biological air filter.
4. Remove the rear chassis cover plate by removing the 3 screws across the
top and 3 screws across the bottom, and lifting the cover plate.
5. Carefully cut the cable ties holding the auxiliary heater thermocouple leads
and disconnect from the Auxiliary Heater Controller Board.
6. Disconnect copper tubing connecting
the auxiliary heater thermocouple to the
auxiliary heater.