STATIM 7000 Service Manual – Conductivity Calibration
54
9. Select “G”, Water conductivity circuit Gain (flashing value on the display), by pressing
Rubber and Plastic Key
10. Adjust G.GG value so that the conductivity in ADC counts (NNN) shows 186 ±1 count.
11. Press Stop key to exit Water conductivity mode and save
displayed
thresholds “HH.H”,
“LL.L” and “G.GG” and enter normal mode of operation, “Select cycle” screen.
Note:
Keypad functions in Conductivity Setup screen
:
Unwrapped Key:
increment current field (the flashing value on the display)
Wrapped Key:
decrement current field (the flashing value on the display)
Rubber and Plastics:
move to next field
Stop key:
exit