STATIM 7000 Service Manual –
Chamber Thermocouples
29
13.2 Chamber Thermocouple (Outlet)
1. Power
OFF
the unit, unplug it and remove the cover. See Cover. NOTE:
The steam generator may be hot if the unit has been operating.
2. Disconnect the fan connectors.
3. Disconnect the inlet and outlet tubes on the biological air filter.
4. Remove the rear chassis cover plate by removing the 3 screws across the
top and 3 screws across the bottom, and lifting the cover plate.
5. Remove the 2 screws that attach the chamber solenoid valve bracket to
the chassis.
6. Remove 2 screws on the under side of the unit that hold solenoid housing
to the chassis.
7. Disconnect the compression nut on
the copper tubing to the auxiliary
heater thermocouple, remove the
screws holding the chamber
solenoid valve base to the chassis
and slide the solenoid with the
copper tubing attached, to the left.