STATIM 7000 Service Manual - Pump
33
14. Pump
If you find a problem with the pump, follow these steps:
1.
Power
OFF
the unit, unplug it and remove the cover. See Cover.
2.
Disconnect the fan connectors.
3.
Disconnect the inlet and outlet tubes on the biological air filter.
4.
Remove the rear chassis cover plate by removing the 3 screws across the
top and 3 screws across the bottom, and lifting the cover plate.
5.
Check the inline water filter. Replace it if necessary.
6.
Cut any cable ties that hold the inlet tube to the pump inlet fitting.
7.
Disconnect the rubber tube
from the pump inlet fitting.
8.
Soak the tube in a mild solution suitable for removing scale or mineral
deposits (e.g., vinegar) until any deposits have been dissolved. Rinse the
tube with clean water.
9.
Disconnect wires 5 and 6 from the Controller Board terminal block.