STATIM 7000 Service Manual - Cassette
2
Replacement seals are available from SciCan (p/n 01-110295S Lid Seal).
When the cassette seal has undergone 1000 cycles, an early message on the
LCD will display SEAL LIFE WARNING - 100 CYCLES REMAINING, with a
warning beep. The seals should be changed immediately. After 1100 cycles, your
unit will stop operating and a message will be displayed SEAL LIFE EXPIRED,
REPLACE. Replacement seals are available from your SciCan dealer.
To change the cassette lid seal, follow these steps:
1. Place the cassette lid and the new seal on a clean work surface.
2. Examine the position of the old seal in the cassette lid and arrange the
new seal in the same orientation, next to the lid.
3. Remove the old seal and discard.
4. Clean any residue out of the seal channel and flush out the channel with
distilled water.
5. Lubricate the new seal with the liquid seal lubricant provided.
6. Align the RFID tag in the new seal with the holes in the lid.
7. Insert the rounded edge of the seal under the round lip of the lid.
NOTE: When inserting the seal, seven round nibs should be visible. The nibs
should fit flush with the lid’s outer surface. Ensure the seal is completely inserted.
Feel around the periphery to ensure the seal is securely seated.
To change the cassette tray seals (O-Rings), follow these steps:
1. Remove the Perforated Rack (01-110294S)
2. Remove the Tray exhaust duct (01-110297S)
3. Remove the Steam Deflector (01-110293S)
4. Unscrew the first coupling (Exhaust Coupling, 01-110292S) and carefully
remove the two O-Ring seals.
5. Find the matching seals from the kit and install them on to the port valve.
6. Secure the first coupling to the cassette ensuring the metal spring is on
the outer wall of the cassette.