STATIM 7000 Service Manual - Cassette
3
7. Unscrew the second coupling (Inlet Coupling, 01-110291S) and carefully
remove the two O-Ring seals.
8. Find the matching seals from the kit and install them on to the port valve.
9. Secure the second coupling to the cassette ensuring the metal spring is
on the outer wall of the cassette.
10. Replace the Tray Exhaust duct, steam deflector, and perforated rack.
NOTE: During a cycle, steam may appear between the lid and the tray. If this
persists, remove the cassette and check that the seal is correctly installed.
Be careful. The metal parts may be hot, and the cassette may contain hot steam.
For the unit to function effectively, the steam deflector and perforated rack MUST
be placed back in the cassette in the proper position.
1.2
Replacing the Cassette Handle
Cracked or otherwise broken Statim 7000 cassette handles can be replaced by
following these steps:
1. Remove the two screws on the underside of each piece of the handle.
2. Remove the handle.
3. Separating the handle from the cassette may require pushing down
forcefully on the handle.
4. Replace and re-attach.