STATIM 7000 Service Manual –
Chamber Solenoid Valve
22
11. Chamber Solenoid Valve
Chamber Solenoid Valve
To remove the Chamber Solenoid (the
solenoid on
left
when facing the back of
the unit), follow these steps:
1.
Power
OFF
the unit, unplug it and
remove the cover. See Cover.
NOTE: The steam generator may
be hot if the unit has been
operating.
2.
Disconnect the fan connectors.
3.
Disconnect the inlet and outlet tubes on the biological air filter.
4.
Remove the rear chassis cover plate by removing the 3 screws across the
top and 3 screws across the bottom, and lifting the cover plate.
5.
Carefully cut the cable ties holding the chamber solenoid leads to the
Auxiliary Heater Controller Board terminal block (wires 15 and 16).