STATIM 7000 Service Manual – Auxiliary Heater
36
15. Auxiliary Heater
To remove the steam generator, follow these steps:
1. Power
OFF
the unit, unplug it and remove the cover. See Cover. NOTE:
The steam generator may be hot if the unit has been operating.
2. Disconnect the fan connectors.
3. Disconnect the inlet and outlet tubes on the biological air filter.
4. Remove the rear chassis cover plate by removing the 3 screws across the
top and 3 screws across the bottom, and lifting the cover plate.
5. Disconnect the auxiliary heater thermocouple from the auxiliary heater
board.
6. Disconnect both P9 wires one from the auxiliary board and one from the
solid state relay. NOTE: the white wire running from the auxiliary board to
the solid state relay should also be replaced when re-installing.
7. Cut the tie wraps holding the auxiliary heater thermocouple.
8. Disconnect the copper tubing connecting the steam generator to the
auxiliary heater.