12 Validation Solenoid Valve
To remove the Validation Solenoid
(solenoid on
right
when facing the back
of the unit), follow these steps.
1. Power
OFF
the unit and remove the
cover. See Cover. NOTE: The
steam generator may be hot if the
unit has been operating.
2. Disconnect the fan connectors.
3. Disconnect the inlet and outlet tubes on the biological air filter.
4. Remove the rear chassis cover plate by removing the 3 screws across the
top and 3 screws across the bottom, and lifting the cover plate.
5. Carefully cut the cable ties holding the validation solenoid leads and
disconnect the validation solenoid from the main Controller Board terminal
block (wires 7 and 8).
6. Disconnect compression nut on the
copper tube at the base of the
solenoid.