STATIM 7000 Service Manual - Pump
35
Pump test bottle set up
15.
Insert the filter and reconnect the inlet fitting to the pump. Ensure that the
rubber seal is in place. Finger tighten the fitting, and then tighten one half
turn with a wrench.
16.
Remove the compression nut from
the tube that connects the pump to
the steam generator.
17.
Connect to the pump test bottle.
18.
Ensure there is ample water in the
service kit reservoir.
19.
Power the unit
ON
, and activate the
pump using the key pad…need more instructions.
20.
Run a sterilization cycle. Verify that the cycle completes normally and
there are no leaks.
21.
Reverse removal instructions to replace.