STATIM 7000 Service Manual – Steam Gen. Thermal Fuse
19
8. Disconnect, at both ends, the
copper tubing connecting the
steam generator to the auxiliary
heater.
9. Remove the two screws from
the steam generator bracket.
NOTE: The steam generator is
still attached to the Controller
board by the thermocouple
lead.
10. Carefully cut the two cable ties closest to the steam generator that hold
the thermocouple leads together.
11. Gently lift and turn the steam
generator assembly onto one
side to expose the bottom of the
assembly.
Be careful not to
stress the
thermocouple leads.
(Min. bend radius - 3/16 inch / 5
mm).