75/90
User manual SCAIME:
NU-Soft-eNod4F-E-0716_235704-C.doc
Name
Modbus
address
CANopen®
Index/sub-
index
Profibus cyclic
IN/OUT
Acyclic DPV1
slot/index
Type Access
Cut-off frequency of totalization
flow rate (x100)
0x0A51
0x4001/ 0x06
/
0x0B / 0x02
Uint
RW
Time unit of totalization flow rate
0x0A52
0x3700/
0x04
/
0x0B / 0x03
Uint
RW
Weight unit of totalization flow
rate
0x0A53
0x3700/
0x05
/
0x0B / 0x04
Ulong
RW
Dynamic zero acquisition time
0x005B
0x4708 / 0x00
R 0x02B0
W 0x02B1
0x07 / 0x1E
Uint
RW
15.2
Settings description
•
Instant flowrate:
loss in weight during a period of time expressed in
weight unit
per
flow rate time unit
.
•
Instant flow rate determination time:
time in seconds used for
instant flow rate
determination.
•
Instant flow rate correction factor:
correction factor coefficient expressed in 10
-6
, so 1000000 corresponds
to a correction factor coefficient of 1.00000 applied on the
instant flow rate
.
•
Average flow rate:
flow rate output of moving average filter on successive values of instant flowrate.
•
Nominal flow rate:
target of flow rate expressed in
weight unit
per
flow rate time unit
.
•
Dosing weight deviation:
weight deviation between the totalization of weight during 1 second and what
should have been totalized at nominal flowrate.
•
Flow rate control output:
this is the flow rate to control the extraction device, expressed in
weight unit
per
flow rate time unit
. It can take either a fixed value of flow rate, or the output of PID controller, if enabled,
or a variable flowrate function of weight value in the hopper (or silo) during refilling phase (in smart
refilling mode).
•
Control output value:
control value of the analog output expressed in 0.01% of maximum output current
or voltage.
•
Refilling low level:
gross weight level below which a refilling has to operate if the
auto refilling
option is
activated, cf § refilling.
•
Refilling high level:
gross weight level above which a refilling has to stop when there is no
refilling inflight
value
.
•
Refilling inflight value:
refilling stops when the gross weight will reach
refilling high level
-
refilling inflight
value
.