Computer-Controlled Operation of the Draft Shield
Control Commands for the Draft Shield on Models MC 210 S (-0CE), MC 210 P (-0CE),
MC 410 S (-0CE) and MC 21S
Open draft shield
to approx. 170°
Close draft shield
Open draft shield as far as stored
position (see also part 1, “Opening and
Closing the Draft Shield”)
Poll draft shield status
Control Commands for the Draft Shield on Models MC 5 (-0CE) and SC 2 (-0CE)
Open draft shield toward the left
to approx. 100°
Close draft shield
Open draft shield as far as stored position
(see also part 1, “Opening and Closing
the Draft Shield”)
Open draft shield toward the right to approx.
100° (absolute position approx. 260°)
Poll draft shield status
Store draft shield position xxx in degrees
and the direction in which it was
or
opened, “left”
Store draft shield position xxx in degrees
and the direction in which it was
opened, “right”
Each control command with the lower-case letter “w” must be terminated
by an underline (ASCII = 5F Hex).
Draft Shield Status
Polling the draft shield status is done to control the motor-driven draft shield door.
The draft shield status will not be indicated until all movement of the draft shield
door has completely stopped (draft shield motor “off”).
ESC t xxx _ CR LF ESC f6 _ CR LF
ESC t xxx _ CR LF ESC f5 _ CR LF
ESC
w
0
_
CR
LF
ESC
w
4
_
CR
LF
ESC
w
3
_
CR
LF
ESC
w
2
_
CR
LF
ESC
w
1
_
CR
LF
ESC
w
0
_
CR
LF
ESC
w
3
_
CR
LF
ESC
w
2
_
CR
LF
ESC
w
1
_
CR
LF
4–10
Содержание MC 210 S
Страница 4: ...General View of the Balances MC 210 S MC 210 P and MC 410 S 0CE 1 0...
Страница 6: ...General View of the Balances MC 5 0CE and SC 2 0CE 1 2 34 35 36 10 7 9 8 37...
Страница 46: ......
Страница 78: ......
Страница 160: ......
Страница 166: ...5 6 Z 14 755 320 34 10 93...
Страница 170: ......
Страница 174: ...6 4...
Страница 178: ......
Страница 179: ......