SEPOR, INC
718 N FRIES AVE.
WILMINGTON, CA 90744
310 830 6601
Fax: 310 830 9336
[email protected]
OPERATING MANUAL
Pulp Density Scale
Страница 1: ...SEPOR INC 718NFRIESAVE WILMINGTON CA90744 3108306601 Fax 3108309336 info sepor com OPERATING MANUAL Pulp Density Scale ...
Страница 2: ... marked for determining the specific gravity of dry solids is used for this purpose A sample of material for determination should be dry and preferably all 10 mesh 100 mesh Hang the dry empty bucket on the pulp density scale then pour in the sample of dry material until the pointer indicates 1000 grams 1 000 kilograms on the outer circle of the dial At this point the bucket will contain one kilogr...
Страница 3: ...r to your operation any of the interchangeable dial faces may be installed as follows 1 Remove the scale laying it flat on a table 2 Loosen the thumbscrews on the sides of the scale and lift off the clear cover and retaining ring 3 Remove the two screws and washers near the center of the dial Do not remove or bend the pointer 4 With pointer pointing directly at your body take the selected dial in ...
Страница 4: ...able Dial Faces are supplied with the scale as follows Dial No SG Solids Range 1 2 2 2 6 2 8 3 2 3 6 4 0 4 4 2 1 2 1 3 1 4 1 5 1 6 1 7 1 8 3 1 7 1 8 1 9 2 0 2 1 2 2 2 3 4 2 2 2 3 2 4 2 5 2 6 2 7 2 8 5 2 6 2 7 2 8 2 9 3 0 3 1 3 2 6 3 0 3 1 3 2 3 3 3 4 3 5 3 6 7 3 4 3 5 3 6 3 7 3 8 3 9 4 0 8 3 8 3 9 4 0 4 1 4 2 4 3 4 4 9 4 2 4 3 4 4 4 5 4 6 4 7 4 8 10 4 6 4 8 5 0 5 2 5 4 5 6 5 8 11 5 6 5 8 6 0 6 2 6...