Practical Example: Dry Weight Determination with Tare Weighing
Menu code settings used in the example:
Function
Code
For MC 210 S,
For
MC 210 P,
MC 5 (-0CE),
MC 410 S (-0CE) SC 2 (-0CE)
and MC 21 S
Differential
weighing and
backweighing
2 1 6
2 1 6
Combined
weighing mode 3 7 3*
3 6 3*
Automatic
data output
7 1 2
7 1 2
With data
ID code
7 2 2
7 2 2
2 decimal
places for the
results in percent 3 6 3*
3 5 3
With tare
weighing
3 8 2
3 7 2*
1st calculated
value: residue
in percent
3 9 2
3 8 2
2nd calculated
value: residue
in wt. unit
3 10 1
3 9 1
Record/printout
with net initial
sample weight
7 5 3*
7 5 3*
Record/printout
with residue
in wt. unit
and percent
7 6 4
7 6 4
Application:
Determination of the dry weight (solids) in percent of samples over
a relatively long period, e.g., using a drying oven. The containers for
the samples are consecutively numbered and placed in the correct
sequence. In this particular application, let’s suppose that your in-
house standard operating procedure does not allow backweighing
to be done immediately after weighing-in a sample.
* = factory setting
As alternatives, you can select
these codes:
Different number
of decimals
3 6 x 3 5 x
Without tare
weighing
3 8 1 3 7 1*
Different calcu-
lated value
3 9 x 3 8 x
Different calcu-
lated value
3 10 x 3 9 x
Other record
printout/output
7 5 x 7 5 x
Other record
printout/output
7 6 x 7 6 x
3–25
Содержание MC 210 S
Страница 4: ...General View of the Balances MC 210 S MC 210 P and MC 410 S 0CE 1 0...
Страница 6: ...General View of the Balances MC 5 0CE and SC 2 0CE 1 2 34 35 36 10 7 9 8 37...
Страница 46: ......
Страница 78: ......
Страница 160: ......
Страница 166: ...5 6 Z 14 755 320 34 10 93...
Страница 170: ......
Страница 174: ...6 4...
Страница 178: ......
Страница 179: ......