Printout/Record for Calibration and Linearization Functions (menu code 8 10 2)
A printout or record is generated only at the end of the following functions:
– all calibration operations
– all linearization operations
– air density determination (see part 3, “Application Programs”)
The printout of the record can have the following lines:
--------------------
MC1 -
Sartorius
: Balance family and manufacturer
Model
MC5
: Balance model
S/N
030800046
: Serial no. of the balance
Id
4-32-1
: ID no. (e.g., for identifying
a workstation/operator)
--------------------
Date : 30-Jul-95
: Current date
Start:
10:05:30
: Time at which the application started
Cal. :
Test
: Calibration mode (in this case, “calibration”)
Diff.:- 0.000013 g
: Data generated during the calibration test
Cal. :
Intern
: Calibration mode (in this case,
“internal calibration”)
Stat.:
Complete
: Status comment for calibration or
linearization functions
End
:
10:05:45
: End of application
Name :
: Field for signature of the operator responsible
--------------------
Set. :
5.000000 g
: Calibration weight (only for “external
calibration”)
Set. :
1.175 kg/
: Air density
Cal. :
Air
2–26