Procedures for the Combined Weighing Mode
Tare Weighing
Symbol displayed:
Net
Key identified by:
●
-
F
Press
F
identified by “
●
” to accept and store a weight as the tare value. The tare
weight will be stored only if the balance has stabilized. At this point, the balance
display is zeroed. After tare weighing, the net weight of a sample is determined.
The ID code of
F
in the display changes from “
●
” to “STO.” Afterwards, weighing
can be done starting from “zero.”
The tare weighing mode can be turned off by setting the appropriate menu code:
Tare weighing
Code
For MC 210 S,
For MC 5 (-0CE),
MC 210 P,
SC 2 (-0CE)
MC 410 S (-0CE)
and MC 21 S
Off
3
8
1*
3
7
1
On
3
8
2
3
7
2*
Weighing-in
Key identified by: STO -
F
The stable net weight is stored when you press the
F
key identified on the display
by “STO.” When the weight is stored, the next memory location ID is automatically
selected.
Before you can proceed from weighing-in to backweighing, the appropriate
memory location ID number must be selected using the numeric keys and
F
.
* = factory setting