When data are output without decimals, the decimal point is suppressed
(except when a certain display mode is selected).
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16
Data output example: +1501.117 mg
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16
Characters:
1st
Plus or minus sign or space
2nd –10th
Weight with a decimal point; leading zeros = space
11th
Space
12th –14th
Unit symbol or space
15th
Carriage return
16th
Line feed
If the weighing system has not stabilized, no unit symbol is output.
For the display format parameter “last numeral off or only at stability” (blank last
numeral by a rounding factor of 10 or display last numeral when the readout
is stable), the 10th character in the string is a space. The reason: the last numeral
without a decimal point is not blanked. Instead, it is given a definite zero,
otherwise the display would go completely blank when the balance is unloaded.
In this case, you will obtain the standard display format.
* = space, E = unit
+
*
1
5
0
1
.
1
1
7
*
m
g
*
CR
LF
+
*
*
*
*
*
*
– – – – – – – – – – – – – – – – – – – – – – – –
*
*
*
*
* 10
7
10
6
10
5
10
4
10
3
10
2
10
1
10
0
*
CR
LF
– – – – – – – – – – – – – – – – – – – – – – – –
E
E
E
–
0
0
0
0
0
0
0
4–4
Содержание MC 210 S
Страница 4: ...General View of the Balances MC 210 S MC 210 P and MC 410 S 0CE 1 0...
Страница 6: ...General View of the Balances MC 5 0CE and SC 2 0CE 1 2 34 35 36 10 7 9 8 37...
Страница 46: ......
Страница 78: ......
Страница 160: ......
Страница 166: ...5 6 Z 14 755 320 34 10 93...
Страница 170: ......
Страница 174: ...6 4...
Страница 178: ......
Страница 179: ......