Step/Key ID/Key
Readout
Data Output
Enter memory location/contianer
ID, e.g.,
10
and store using
F
10
Zero the balance using
t
0.000 mg
Place container on pan;
+ 355.916 mg
store weight using
●
–
F
0.000 mg
Net
Add sample to container;
+ 780.352 mg
Net
store weight using STO –
F
+1136.268 mg
11
Remove container with sample
0.000 mg
Load next container;
+ 375.235 mg
store weight using
●
-
F
0.000 mg
Net
Add sample to container
+ 889.715 mg
Net
store weight using STO -
F
+1264.950 mg
Store weights of other
12
containers and weigh-in
further samples
Treat samples (e.g., dry samples)
Enter 1 st memory location/
container ID
10
and
10
press to confirm
●
-
F
Zero the balance using
t
0.000 mg
Place corresponding con 934.696 mg
on pan; backweigh using
+
74.17 %
g
Net
Num
10
RE S -
F
Date
9-Jul-95
Time
10:34:54
Net
+
780.352 mg
Res
+
578.780 mg
Res
+
74.17 %
If necessary, press
w
+ 578.780 mg
Net
to display 2nd calculated value
Repeat procedure for all
samples RE S v -
F
11
3–26