Control Lines
For classification or filling in the “over/under checkweighing” application program,
you can use the voltage levels of four data output port lines to control an external
on-line display or control instrument. To activate the control lines, set code 8 7 1
in the balance operating menu.
The voltage levels of the data output ports will change according to various patterns,
depending on the reference weight and on the lower and upper weight limits.
Lower limit
Reference weight
Upper limit
Control output
Pin
Voltage
Voltage
level
level
1
Low
High
16
2
High
High
17
3
High
Low
18
4
High
High
19
Control output 1: lighter
Control output 2: equal
Control output 3: heavier
Control output 4: “set” (target)
4–18
Содержание MC 210 S
Страница 4: ...General View of the Balances MC 210 S MC 210 P and MC 410 S 0CE 1 0...
Страница 6: ...General View of the Balances MC 5 0CE and SC 2 0CE 1 2 34 35 36 10 7 9 8 37...
Страница 46: ......
Страница 78: ......
Страница 160: ......
Страница 166: ...5 6 Z 14 755 320 34 10 93...
Страница 170: ......
Страница 174: ...6 4...
Страница 178: ......
Страница 179: ......