7-30
ECL-210 Operation Manual
The fine tuning amount, which is auto-set independently by setting wavelength, cannot be
saved to unit memory by STORE command.
NOTE
7.16. Plug and Play
When units are exchanged or added in old type system, the stable system must be turned off and
units are exchanged or added, and then system must be turned on again. However the units can be
exchanged and added as the system is on in this system. Turning the system on again is not
necessary. The warm up of the unit is only required.
7.16.1. Unit Insertion
Here shows the procedure of inserting a unit.
1) Insert unit along the guide of MLS slot.
2) Insert the unit to the bottom and firmly fix the unit fixing screw on the left top of unit by
rotating.
3) MLS confirms the slot in which unit is inserted to MLS.
4) After MLS self check, the unit LED is lit if there is no problem.
5) MLS starts supplying power source and communication control to the new unit.
NOTE
NOTE
Make warm up for several minutes of the new unit.
In inserting some new units, insert the next unit after the first unit insertion sequence finished
completely. Do not insert units at the same time. It may cause damage.