7-16
ECL-210 Operation Manual
NOTE
NOTE
NOTE
The direct angle sensor is attached to OBPF. The angle sensor enables the actuator to feed-
back control. There may be cases of little overshoot and undershoot.
The monitor signals of thermistor, attached to FLF, are temperature-controlled precisely by
PI control. The FLF control takes time, and the overshoot and undershoot may be repeated.
In that case, wait a little (nearly 1 minute) to be stable adequately, though it seems wave-
length to be stable.
The present value of wavelength is calculated from OBPF sensor signals and FLF sensor
signals. The excellent FLF characteristics for selecting wavelength enable precise accuracy
on the display. The accuracy is approximately
±
3GHz.