EXTERNAL CAVITY TUNABLE LD LIGHT SOURCE ECL-210
6-3
6.4. Heat Dissipation Environment
The fan for heat dissipation is located in back of MLS. Please secure an opening 5cm or more
from a fan, and air should flow into it from the slit of the bottom front part. If the slit is closed, or
the air doesn't espace from the bottom properly, the equipment generates heat and it may not
operate stably. It also causes the same when the blank cover is not inserted in the empty slot.
When there is an empty slot, please be sure to fix a blank cover firmly.
A blank cover has the effect which keeps the flow of the cooling inside a main part.
Since the rubber legs are for securing a slit, please use rubber legs, without removing them
from this equipment.
NOTE
Slit
NOTE