7-6
ECL-210 Operation Manual
:
:
:
TYPE
OPT
S / N
ECL-210
PMF
001
LF.IN
RF.IN
7.2.5. Turning ECL Power ON
MLS confirms the insertion state of each unit, and supplies power sources to the slots in-
serted. The supplied units start up independently and those are communicated with MLS and
controlled.
When the system is turned ON, the following actions (Sequence) are carried out until the
system is activated.
1) Initialize optical systems and control circuits.
2) If the self-checking is okay, LEDs are lit.
3) Cancel the LD protection circuit
4) Supply the power for driving LD from MLS (cancel the protection circuit in MLS)
5) Start communication with MLS when the communication control is ready.
6) Set each parameters, and start up the conditions memorized