EXTERNAL CAVITY TUNABLE LD LIGHT SOURCE ECL-210
7-17
7.10. Setting Optical Output Power
NOTE
Here explains the setting optical power on the assumption that LD is driving. (Refer 7.
8. Driving LD and Shutting LD off)
Setting optical power is available only with dBm unit. It is same as in communication control.
NOTE
7.10.1. APC Function
It is necessary to operate the machine in APC mode to set optical output power. Optical output
cannot be set in ACC mode. Be sure that the mode is in APC. If ACC, change the mode to
APC. If the LED is not lit in the [LD.ON] line of the Matrix LD to operate, the mode is ACC
and it is required to change to APC.
7.10.1.1. Procedure of Setting for ACC Operation
1) Select the unit to operate on MLS. (Refer 7.4. Selecting Units in MLS)
2) Use [MODE] key to set the MDOE APC.