3) Press [ON/OFF] key and the LED on the [APC.ON] line is lit.
7.10.1.2. Setting Optical Output Power
Here explains the setting optical power on the assumption that LD is driving and that the
MODE is APC. (Refer 7.8. Driving LD and Shutting LD off)
1) Select the unit to operate on MLS. (Refer 7.4. Selecting Units in MLS)
2) Use [MODE] key to set the MDOE either APC, COH, or POWER.
3) Use the shuttle switch to set optical output power. (Refer 7.7. Shuttle switch)
4) MLS controls the set optical output power.
If the system has the ATT option, the control method is different. The optical output power is
originally controlled by controlling injection current. But when to set low optical output
power, if the injection current is controlled, it causes worse S/N of signals. In the case of the
built-in ATT option, the ATT processes the rough adjustment, and current does the subtle
fluctuation and adjustment. In that case, it does not worse S/N of signals.