EXTERNAL CAVITY TUNABLE LD LIGHT SOURCE ECL-210
7-29
The delimiters are CR, LF, CR+LF, and EOI without characters. The address can be set 0-30.
It moves by rotating the shuttle.
NOTE
7.15. Parameter Memory of Each Unit
After setting and changing the parameters of each unit from MLS, the state of each unit can be
saved to memory. When you turn on the system again, the current situation appears again.
1) Select the unit to operate on MLS. (Refer 7.4. Selecting Units in MLS)
2) Press [STORE] key.
3) The STORE command is sent to the unit selected by MLS.
Memorized parameters are injection current set value for LD, temperature set value, APC
optical power set value, APC/ACC, COH control ON/OFF, COH control amount, and wave-
length (Fine tuning set value)