79
Press [Mkr -> Stop Freq] and set the stop frequency equal to the marker frequency.
e) Press
【
Mkr
】
→ [Off] to turn off the frequency of the current active marker or difference marker.
f)
Press
【
Mkr
】
→ [All Off] to turn off the frequency of all active or difference markers.
g) Press
【
Mkr
】
→ [Counter Off On] to open the counter function of the currently active marker. The
reading of the counter is not affected by the frequency offset function. To make the counting function
work properly, SPAN/RBW must be less than 500. The counting function can only measure
continuous wave signals or discrete spectral components, and the signal amplitude is greater than
-50dBm, and the signal amplitude must be greater than the noise level by 30dB.
NOTE:
In order to accurately measure the frequency when performing the
frequency counting function, the signal generator and Analyzer to be
tested should be shared with the time base!
Figure 5-6 Marker Function
Figure
!
Содержание S5105 Series
Страница 1: ...S5105 Series Microwave Multifunctional Analyzer User Manual Saluki Technology Inc...
Страница 12: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 6 Part I Operation instructions...
Страница 57: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 51 Figure 4 11 Menu structure of sweep...
Страница 171: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 165 Cal Cont Complete the remaining calibration steps...
Страница 178: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 172 Part II Technical Instructions...
Страница 181: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 175 Part III Maintenance Instructions...