signal frequency one IF higher than the LO. In this case, one of the signal is called the mirror frequency
of the other. Each frequency of the LO has a mirror frequency of corresponding input signal, with the
difference of two IFs between the signal and frequency of the mirror frequency.
Test image, multiple and out-of-band responses on all frequency bands. The signal is applied to the input
port of the spectrum analyzer for reference amplitude measurement. The composite signal generator is
then tuned to a frequency that causes a mirror, multiple or out-of-band response, and the amplitude
displayed on the spectrum analyzer is measured and recorded.
a)
Testing equipment
Composite Signal Generator ..................................................................................................1464B/C
Test Cables and Adapters
b)
Testing steps
10MHz IN
Adapter
10 MHz o
Figure 5-38 Block Diagram for Testing Mirror, Multiple, and Out-of-Band Response
1) Connect the test equipment according to Figure 5-38. The microwave analyzer is used to
provide reference frequencies for the composite signal generator.
2) Set the 1464B/C frequency to 2GHz and the power to -10dBm.
3) Set the spectrum analyzer as follows:
Center frequency ...................................................................................................................2GHz
Span......................................................................................................................................10kHz
Reference Level................................................................................................................ -10 dBm
Resolution bandwidth ............................................................................................................ 10Hz
Video bandwidth .................................................................................................................... 10Hz
4) Adjust the output power level of composite signal generator so that the signal peak is close to
the reference level of the spectrum analyzer.
5) Press [Peak] ->
【
Mkr
】
-> [Difference mode] on the spectrum analyzer.
6) Set 1464B/C to each frequency point of mirroring, multiple and out-of-band response
corresponding to 2GHz listed in the test item of
S5105D/E/F Microwave Analyzer Record Form
,
set the reference level of the microwave analyzer to -40dBm, Press [Peak] , and record the
difference marker amplitude value as the response amplitude in the corresponding test item of
the
S5105 D/E/F Microwave Analyzer Record Form
7) For the mirror, multiple and out-of-band responses of the residual frequency points listed in the
S5105 D/E/F Microwave Analyzer Record Form
, refer to the settings in steps 2 - 6 until the test
of all frequencies is completed.
9
Residual Response
Description: it refers to the discrete response displayed on the display when the spectrum analyzer has
not connected with any input signal.
Содержание S5105 Series
Страница 1: ...S5105 Series Microwave Multifunctional Analyzer User Manual Saluki Technology Inc...
Страница 12: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 6 Part I Operation instructions...
Страница 57: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 51 Figure 4 11 Menu structure of sweep...
Страница 171: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 165 Cal Cont Complete the remaining calibration steps...
Страница 178: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 172 Part II Technical Instructions...
Страница 181: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 175 Part III Maintenance Instructions...