[SPAN/RBW 100], [RBW/VBW 100], and [Average Off On ] .
[Ave Off On]
Turn on/off averaging function, and the power values are continuously averaged to obtain an average of
multiple power values. If the average is off, “Average Off” will be displayed on the left side of the screen,
and “Average On” will be displayed when opening. The average number of times is set from 1 to 1,000.
Section 6 Technical Specifications
The S5105D/E/F microwave analyzer is kept for two hours at ambient temperature, and preheating for
15 minutes, the following performance characteristics should be met in the power monitoring mode.
a)
Frequency range:
100kHz - 18GHz (S5105D)
100kHz - 26.5GHz (S5105E)
100kHz - 40GHz (S5105F)
b)
Power range:
-60dBm to +20dBm (S5105D/E)
-50dBm to +20dBm (S5105F)
Содержание S5105 Series
Страница 1: ...S5105 Series Microwave Multifunctional Analyzer User Manual Saluki Technology Inc...
Страница 12: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 6 Part I Operation instructions...
Страница 57: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 51 Figure 4 11 Menu structure of sweep...
Страница 171: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 165 Cal Cont Complete the remaining calibration steps...
Страница 178: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 172 Part II Technical Instructions...
Страница 181: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 175 Part III Maintenance Instructions...