Menu Description
【
Frep
】
Soft key menu related to frequency pops up, including [Start Freq], [Stop Freq], [Center Freq] and [Span].
[Start Freq]
Activate the start frequency and set the instrument to the start frequency/stop frequency mode. Adjust
the start frequency with a figure key, a step key or a rotary knob. If the start frequency set is greater than
the stop frequency, the start frequency is automatically set to equal the stop frequency.
[Stop Freq]
Activate the stop frequency, and set the instrument to start/stop frequency mode. Adjust the stop
frequency with a figure key, a step key or a rotary knob. If the stop frequency set is smaller than the start
frequency, the stop frequency is automatically set to equal the start frequency.
[Center Freq]
Activate the center frequency, and set the instrument to center frequency/span mode. Adjust the center
frequency with a figure key, a step key or a rotary knob. If the center frequency set is not in harmony with
the current span, the span will be automatically adjusted to the best value corresponding to the center
frequency.
[Span]
Activate the span, and set the instrument to center frequency/span mode. Adjust the span with a figure
key, a step key or a rotary knob.
【
Amplitude
】
[Top]
Set the maximum scale in the top position of the vertical part of the display grid within the range of
-499.9dB to +500dB.
[Bottom]
Set the maximum scale in the top position of the vertical part of the display grid within the range of
-500dB to +499.9dB.
[Auto Scale]
Display the trace in the scale most suitable for the screen, which is only valid in Cartesian coordinates.
[Default scale]
All options in the Cartesian coordinate format for the amplitude setting are the default values at power
on.
[Ref Value]
Set the reference level, which is only valid in Cartesian coordinates.
[Ref Pos]
Set the reference position, which is only valid in Cartesian coordinates.
[Scale]
Set the scale/div, which is only valid in Cartesian coordinates.
[Output Power High>]
Set the output power to high, low or manually set.
[High]
Set the output power to high.
[Low]
Set the output power to low.
Содержание S5105 Series
Страница 1: ...S5105 Series Microwave Multifunctional Analyzer User Manual Saluki Technology Inc...
Страница 12: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 6 Part I Operation instructions...
Страница 57: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 51 Figure 4 11 Menu structure of sweep...
Страница 171: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 165 Cal Cont Complete the remaining calibration steps...
Страница 178: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 172 Part II Technical Instructions...
Страница 181: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 175 Part III Maintenance Instructions...