![Saluki S5105 Series Скачать руководство пользователя страница 45](http://html1.mh-extra.com/html/saluki/s5105-series/s5105-series_user-manual_3686697045.webp)
Tel: 886.909 602 109 Email: [email protected]
39
extending the sweep time.
Steps for setting the track smoothing:
Press
【
BW
】
→ [Smooth Off On] and input the smooth aperture with numeric keys.
Reduce IF bandwidth
The microwave analyzer converts the received response signal to a lower intermediate frequency signal
for processing. The bandwidth of the intermediate frequency band pass filter is called the intermediate
frequency bandwidth, and the microwave analyzer supports the variable bandwidth IF filter, wherein the
intermediate frequency bandwidth is set from 100 kHz to 1 Hz and changes in steps of 1, 3, and 10.
Reducing the IF bandwidth can lower the impact of the random noise on the measurements (reducing 10
times of the IF bandwidth can lower the noise base by 10dB), but it can prolong the sweep time. The
effect of IF bandwidth on measurements is shown in Figure 4-2:
10Hz IF bandwidth
3,000Hz IF bandwidth
Figure 4-2 Effect of IF Bandwidth on Measurement Results
Follow the steps below to set the IF bandwidth:
a)
Press
【
BW
】
→ [IF Band]
;
b)
Press the [↑] [↓] key to move the marker bar to change the IF bandwidth value.
NOTE:
Increasing dynamic range can prolong measurement time due to some
settings of the microwave analyzer.
II
Calibration for higher measurement accuracy
1
Process of correcting measurement error
·
Frequency response correction
·
Frequency response and isolation correction
·
Single port reflection correction
·
Full 2 ports correction
2
Error correction type
Several types of error correction are provided, with 1-12 system errors available: Full 2 ports correction
eliminates effectively all the 12 correctable system errors. All the 12 errors are not required for correction
at the time for some measurements. Each correction and usage are described below:
!
Содержание S5105 Series
Страница 1: ...S5105 Series Microwave Multifunctional Analyzer User Manual Saluki Technology Inc...
Страница 12: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 6 Part I Operation instructions...
Страница 57: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 51 Figure 4 11 Menu structure of sweep...
Страница 171: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 165 Cal Cont Complete the remaining calibration steps...
Страница 178: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 172 Part II Technical Instructions...
Страница 181: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 175 Part III Maintenance Instructions...