S5105E Microwave Analyzer Test Results (Continued 4)
No.
Item
Unit
Standard requirement
Results
26
interface
RF interface
/
3.5 mm positive
Communication
Interface
/
USB interface: type A, two
/
USB interface: mini type, one
/
LAN interface: RJ45 type
Frequency
reference
/
Reference input/output, BNC female type
Other interface
/
GPS antenna interface: BNC female type
/
External trigger input interface: BNC female type
27
Safety
/
Dielectric Strength requires AC 1,500V, 10mA/min;
no breakdown, no arcing.
/
Voltage 242 V, leakage current ≤3.5 mA, 1min.
/
The insulation resistance between the power input
end and the casing should not be less than 100 MΩ
under standard atmospheric conditions for testing.
/
The insulation resistance between the power input
end and the casing should not be less than 2 MΩ in
humid environment.
Description
1. “√”indicates normal function or meet the requirements; “X” indicates that the function is abnormal or does
not meet the requirements;
2. "/" indicates that there is no such test item on this device.
Comprehensive judgment: Pass
□
Fail
□
Содержание S5105 Series
Страница 1: ...S5105 Series Microwave Multifunctional Analyzer User Manual Saluki Technology Inc...
Страница 12: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 6 Part I Operation instructions...
Страница 57: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 51 Figure 4 11 Menu structure of sweep...
Страница 171: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 165 Cal Cont Complete the remaining calibration steps...
Страница 178: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 172 Part II Technical Instructions...
Страница 181: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 175 Part III Maintenance Instructions...