Guidance and manufacturer's declaration – electromagnetic immunity
The Ventilator Respina P1 is intended for use in the electromagnetic environment specified below.
The customer or the user of the Ventilator Respina P1 should assure that it is used in such an
environment.
Immunity test
Port
Compliance level
Electromagnetic
environment - guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
Enclosure
±8 kV contact
± 2 kV, ± 4 kV, ± 8
kV, ±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Patient coupling
Signal input/output
parts
Electrical fast
transient/burst
IEC 61000-4-4
Input a.c. power
± 2 kV, 100 kHz
repetition frequency
Mains power quality
should be that of a typical
commercial or hospital
environment.
Signal input/output
parts
± 1 kV
100 kHz repetition
frequency
Surge
IEC 61000-4-5
Input a.c. power
± 0,5 kV, ± 1 kV
Line-to-line
± 0,5 kV, ± 1 kV, ± 2
kV Line-to-ground
Mains power quality
should be that of a typical
commercial or hospital
environment.
Signal input/output
parts
± 2 kV Line-to ground
Voltage dips,
IEC 61000-4-11
Input a.c. power
0 % U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycles
Single phase: at 0°
Voltage interruptions
IEC 61000-4-11
Input a.c. power
0 % U
T
; 250/300 cycle
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
Enclosure
30 A/m
50 Hz or 60 Hz
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to application of test level.