95
NIF
(The maximum negative pressure generated during
inspiration against an occluded airway)
Accuracy: ± (2 cmH20 + 4%)
-60 – 0 cmH2O
P0.1/NIF
0 – 100 %
WOBimp
(Work of breathing)
Accuracy: ± 10% Actual reading
0 – 20 j/l
RCe
(Expiratory time constant)
0 – 99.9 sec
Gas Temp
Accuracy: 0.1
0-100 °C
Real Time Curves
Pressure - Time , Flow - Time , Volume - Time
Pressure over time, Flow over time,
Volume over time
P–V Loop, F–V Loop
Pressure–volume loop, Flow–volume
loop
Performing
Procedures
Inspiration-Hold
Pause the ventilator at end inspiration
Maximum allowable duration shall
be 20 sec
Expiration-Hold
Pause the ventilator in expiration
Maximum allowable duration shall
be 20 sec
Manual Breathing
Delivery a mandatory breath during
the next expiratory breath in all mode
P0.1
The maximal slope of the airway
pressure drop during the first 100 ms
when the airway is occluded.
NIF
The maximum negative pressure
generated during inspiration against
an occluded airway
Suction
This function allows breaths to be
delivered with a 100% O2
concentration, before and after a
tracheal suction maneuver.
Suction support disconnection time:
20 – 120 s; Resolution: 10
O2 100%
Delivers O2 100% for 2 minutes
Standby
Places ventilator in standby mode
Nebulizer
During the nebulization time, the
nebulizer valve opens synchronous to
inspiration to provide pressure for an
external medication nebulizer
Advanced
Features
Dual control in the breath
Switching from Volume to Pressure
control (and vice versa) during the
inspiratory phase of each breath cycle
(In square flow pattern).
Available in VCV and V-SIMV
modes.
Auto Tlow
Automatically determination of Tlow
to terminate expiration at an
adjustable percentage of peak
expiratory flow.
Available in APRV mode.