The nebulizer will also be disabled in the below conditions and “Neb. Not Usable” alarm
message will be displayed (to avoid deviations in flow delivery by the nebulizer):
- If high pressure oxygen source is not available
- If the oxygen pressure is out of the allowed range.
When the nebulizer is enabled, flow and volume waveforms and parameters will be
compensated.
If the nebulizer is activated but the conditions for activating the nebulizer are not provided,
the system immidiately stops the drug nebulization and after the nebulizer limit is removed,
the device continues the remaining time of drug nebulization.
Note:
Since 100% O2 is administered for drug nebulization, the oxygen concentration of delivered
gas to patient may exceed the user setting during the nebulization.
If you press the Standby key, the maneuver will stop. The maneuver will resume if you exit
the system from the standby mode (The nebulization message is displayed during this time).
If user selects “New Patient” the maneuver will be ended.
This maneuver cannot be enabled in the standby mode.
Humidification type
Three options are available for Humidification including:
Humidifier
HME
None
Warning
HME is disposable and should be exchanged for each patient.
Compliance Compensation
Set
Compliance Compensation
to On/Off in this menu.